CVE-2019-25723 PUBLISHED

Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface

Assigner: VulnCheck
Reserved: 02.06.2026 Published: 02.06.2026 Updated: 03.06.2026

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

Metrics

CVSS Vector: CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L
CVSS Score: 6.3

Product Status

Vendor Dräger
Product Perseus A500
Versions Default: unknown
  • affected from 2.00 to VG2.3.1 (incl.)
  • Version 2.01 is affected
  • Version 2.02 is affected
  • Version 2.03 is unaffected

References

Problem Types

  • CWE-1286 Improper Validation of Syntactic Correctness of Input CWE