CVE-2025-15653 PUBLISHED

Dräger Zeus IE Anesthesia Workstation USB Interface Privilege Escalation

Assigner: VulnCheck
Reserved: 02.06.2026 Published: 02.06.2026 Updated: 03.06.2026

Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.

Metrics

CVSS Vector: CVSS:4.0/AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
CVSS Score: 7

Product Status

Vendor Dräger
Product Zeus IE
Versions Default: unknown
  • affected from 0 to 1.0.5 (excl.)
Vendor Dräger
Product Zeus RS C500
Versions Default: unknown
  • affected from 0 to 1.0.9 (excl.)

References

Problem Types

  • CWE-668 Exposure of Resource to Wrong Sphere CWE